Performance Features
Typical Applications
Online monitoring of total nitrogen in industrial wastewater and municipal sewage discharge.
Technical Specifications
| Item | Specification |
| Method Basis | Potassium Persulfate Oxidation - Ultraviolet Spectrophotometric Method |
| Measurement Range | 0-20mg/L; 0-100mg/L (dual-range switchable, extendable) |
| Indication Error | ≤0.5mg/L: ±0.05mg/L; ≥0.5mg/L: no more than ±10% |
| Repeatability | ±3% |
| Stability | No more than ±10% within 24 hours |
| Measurement Cycle | ≤50min |
| Sampling Cycle | Time interval (adjustable arbitrarily from 10 min to 9999 min) and on-the-hour measurement mode |
| Calibration Cycle | Adjustable at any interval from 1 to 99 days and at any time |
| Maintenance Cycle | Generally once a month, about 30 minutes each time |
| Reagent Consumption | Less than 0.5 RMB per sample |
| Output | 4~20mA (2 channels), RS232, RS485 |
| Environmental Requirements | Indoor with adjustable temperature, recommended temperature +5~28℃; Humidity ≤90% (no condensation) |
| Power Supply | AC220±10% V, 50±10% Hz, 5 A |
| Dimension | 800 mm (H) × 430 mm (W) × 300 mm (D) |
| Others | 1. Abnormal alarm and power failure will not lose data; 2. Touch screen display and command input; 3. After abnormal reset and power recovery, the instrument automatically discharges residual reactants inside and resumes work automatically. |
Working Principle
In an alkaline medium at 120℃~124℃, potassium persulfate oxidant is added to oxidize ammonia, ammonium salts, nitrites and most organic nitrogen compounds in the water sample into nitrates. Then, the total nitrogen content is determined by the alkaline potassium persulfate digestion-ultraviolet spectrophotometric method in the form of nitrate nitrogen.